
GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
138 of 157
A.1.1.1 Implementation Specific details
• For all CT Images created by LightSpeed patient birth date element (0010,0030) is
sent, if the operator has entered the details. If the operator has not entered the details
it will be sent as a “Zero length element”
• The length of Accession number element (0008,0050) is a maximum of 16 characters
• Window Center (0028,1050) and Window Width (0028,1051) are sent for all CT
Images created by LightSpeed
• Based on whether contrast was used or not the following applies for CT Images
created by LightSpeed
Mode (0018,0010) (0018,1040)
No Contrast Not sent Not sent
Oral Contrast Oral Contrast Agent name The string “Oral”
IV Contrast IV Contrast agent name The String “IV”
Oral and IV
contrast
Oral contrast agent “ & “
IV Contrast agent
The string “Oral & IV”
• Supported field length for DICOM fields
DICOM Tag Field Name Field Length
( 0008, 0050 ) Accession Number 16 characters
( 0010, 0010 ) Patient Name 32 characters
(configurable to 64)
( 0010, 0020 ) Patient Id 16 characters
(configurable to 64)
A.2 SC Image IOD
A.2.1 SC Image IOD Modules
Entity Name Module Name Reference Usage
Patient Patient A.3.1 M
Study General Study A.3.2 M
Patient Study A.3.3 U
Series General Series A.3.4 M
Equipment General Equipment A.3.5 U
SC Equipment A.5.1 M
Comentários a estes Manuais